Caution
Do not drive the same PXI local bus line with the NI PXI-781xR and another
device simultaneously. Such signal driving can damage both devices. NI is not liable for
any damage resulting from such signal driving.
The NI PXI-781xR local bus lines are compatible only with 3.3 V signaling
LVTTL and LVCMOS levels.
Caution
Do not enable the local bus lines on an adjacent device if the device drives
anything other than 0–3.3 V LVTTL signal levels on the NI PXI-781xR. Enabling the lines
in this way can damage the NI PXI-781xR. NI is not liable for any damage resulting from
enabling such lines.
The left local bus lines from the left peripheral slot of a PXI backplane
(Slot 2) are routed to the star trigger lines of up to 13 other peripheral slots
in a two-segment PXI system. This configuration provides a dedicated,
delay-matched trigger signal between the first peripheral slot and the
other peripheral slots and results in very precise trigger timing signals.
For example, an NI PXI-781xR in Slot 2 can send an independent
trigger signal to each device plugged into Slots <3..15> using the
PXI/LBLSTAR<0..12>. Each device receives its trigger signal on its own
dedicated star trigger line.
Caution
Do not configure the NI PXI-781xR and another device to drive the same physical
star trigger line simultaneously. Such signal driving can damage the NI PXI-781xR and the
other device. NI is not liable for any damage resulting from such signal driving.
Refer to the PXI Hardware Specification Revision 2.1 and PXI Software
Specification Revision 2.1 at
www.pxisa.org
for more information about
PXI triggers.
Switch Settings
Refer to Figure 2-3 for the location of the switches on the NI 781xR. For
normal operation, SW1 is in the OFF position. To prevent a VI stored in
flash memory from loading to the FPGA at power up, move SW1 to the
ON position, as shown in Figure 2-5.
Note