beautypg.com

Table 4-1. input ranges for ni 6232/6233, Analog input ground-reference settings, Analog input ground-reference settings -3 – National Instruments NI 6232 User Manual

Page 30: Analog input, Ground-reference settings

background image

Chapter 4

Analog Input

© National Instruments Corporation

4-3

NI 6232/6233 User Manual

16-bit ADC converts analog inputs into one of 65,536 (= 2

16

) codes—that

is, one of 65,536 possible digital values. These values are spread fairly
evenly across the input range. So, for an input range of –10 V to 10 V, the
voltage of each code of a 16-bit ADC is:

M Series devices use a calibration method that requires some codes
(typically about 5% of the codes) to lie outside of the specified range. This
calibration method improves absolute accuracy, but it increases the nominal
resolution of input ranges by about 5% over what the formula shown above
would indicate.

Choose an input range that matches the expected input range of your signal.
A large input range can accommodate a large signal variation, but reduces
the voltage resolution. Choosing a smaller input range improves the voltage
resolution, but may result in the input signal going out of range.

For more information on programming these settings, refer to the
NI-DAQmx Help or the LabVIEW 8.x Help.

Table 4-1 shows the input ranges and resolutions supported by the
NI 6232/6233 device.

Analog Input Ground-Reference Settings

NI 6232/6233 devices support the analog input ground-reference settings
shown in Table 4-2.

Table 4-1. Input Ranges for NI 6232/6233

Input Range

Nominal Resolution Assuming

5% Over Range

–10 V to 10 V

320

μV

–5 V to 5 V

160

μV

–1 V to 1 V

32

μV

–200 mV to 200 mV

6.4

μV

(10 V – (–10 V))

2

16

= 305 mV

This manual is related to the following products: