beautypg.com

Cables and accessories, Custom cabling, Cables and accessories -4 – National Instruments NI 6232 User Manual

Page 23: Custom cabling -4

background image

Chapter 2

DAQ System Overview

NI 6232/6233 User Manual

2-4

ni.com

Cables and Accessories

NI offers a variety of products to use with NI 6232/6233 devices, including
cables, connector blocks, and other accessories, as follows:

Cables and cable assemblies

Shielded

Unshielded ribbon

Screw terminal connector blocks, shielded and unshielded

RTSI bus cables

For more specific information about these products, refer to

ni.com

.

Refer to the Custom Cabling section of this chapter, the

Field Wiring

Considerations

section of Chapter 4,

Analog Input

, and Appendix A,

Device-Specific Information

, for information on how to select accessories

for your M Series device.

Custom Cabling

NI offers cables and accessories for many applications. However, if you
want to develop your own cable, the following kits can assist you:

TB-37F-37SC—37-pin solder cup terminals, shell with strain relief

TB-37F-37CP—37-pin crimp & poke terminals, shell with strain
relief

Also, adhere to the following guidelines for best results:

For AI signals, use shielded, twisted-pair wires for each AI pair of
differential inputs. Connect the shield for each signal pair to the ground
reference at the source.

Route the analog lines separately from the digital lines.

When using a cable shield, use separate shields for the analog and
digital sections of the cable. Failure to do so results in noise coupling
into the analog signals from transient digital signals.

For more information on the connectors used for DAQ devices, refer to the
KnowledgeBase document, Specifications and Manufacturers for Board
Mating Connectors
, by going to

ni.com/info

and entering the info code

rdspmb

.