EMC Information

CAUTION

The RadiForce series require special precautions regarding EMC and need to be installed, put

into service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.

Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the RadiForce

series. Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If

adjacent or stacked use is necessary, the equipment or system should be observed to verify

normal operation in the configuration in which it will be used.

Guidance and manufacturer’s declaration - electromagnetic emissions

The RadiForce series are intended for use in the electromagnetic environment specified below.

The customer or the user of the RadiForce series should assure that it is used in such an environment.

Emission test

Compliance Electromagnetic environment - guidance

RF emissions

EN55011

Group 1

The RadiForce series use RF energy only for its internal function.

Therefore, its RF emission are very low and are not likely to cause

any interference in nearby electronic equipment.

RF emissions

EN55011

Class B

The RadiForce series are suitable for use in all establishments,

including domestic establishments and those directly connected to

the public low-voltage power supply network that supplies buildings

used for domestic purposes.

Harmonic emissions

EN61000-3-2

Class D

Voltage fluctuations /

flicker emissions

EN61000-3-3

Complies

Guidance and manufacturer’s declaration - electromagnetic immunity

The RadiForce series are intended for use in the electromagnetic environment specified below.

The customer or the user of the RadiForce series should assure that it is used in such an environment.

Immunity test

EN60601

test level

Compliance level

Electromagnetic environment -

guidance

Electrostatic

discharge (ESD)

EN61000-4-2

±6kV contact

±8kV air

±6kV contact

±8kV air

Floors should be wood, concrete

or ceramic tile. If floors are

covered with synthetic material,

the relative humidity should be at

least 30%.

Electrical fast

transient / burst

EN61000-4-4

±2kV

for power supply lines

±1kV

for input/output lines

±2kV

for power supply lines

±1kV

for input/output lines

Mains power quality should be

that of a typical commercial or

hospital environment.

Surge

EN61000-4-5

±1kV differential mode

±2kV common mode

±1kV differential mode

±2kV common mode

Mains power quality should be

that of a typical commercial or

hospital environment.

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines

EN61000-4-11

<5% U

T

, for 0.5 cycle

(>95% dip in U

T

)

40% U

T

, for 5 cycles

(60% dip in U

T

)

70% U

T

, for 25 cycles

(30% dip in U

T

)

<5% U

T

, for 5s

(>95% dip for U

T

)

<5% U

T

, for 0.5 cycle

(>95% dip in U

T

)

40% U

T

, for 5 cycles

(60% dip in U

T

)

70% U

T

, for 25 cycles

(30% dip in U

T

)

<5% U

T

, for 5s

(>95% dip for U

T

)

Mains power quality should be

that of a typical commercial or

hospital environment. If the user

of the RadiForce series requires

continued operation during

power mains interruptions, it is

recommended that RadiForce

series be powered from an

uninterruptible power supply or

battery.

Note: U

T

is the a.c. mains voltage prior to

application of the test level.

Power frequency

(50/60Hz)

magnetic field

EN61000-4-8

3A/m

3A/m

The power frequency magnetic

field should be measured in the

intended installation location to

assure that it is sufficiently low.