Message what it means, Action – Lifescan UltraSmart User Manual
Page 58
49
MESSAGE
WHAT IT MEANS
Error message that
could be caused by
either a used test
strip or a problem
with the meter.
Error message that
indicates that the
blood or control
solution sample was
applied before
“Apply Blood,”
“Alternate Site,” or
“Control Solution”
appeared on the
display.
ACTION
Repeat the test with
a new test strip. Refer
to page 7–8 for test
strip information. If
the error message
persists, contact your
local LifeScan
Customer Service
office.
See back cover
for contact
information.
Repeat the test with
a new test strip.
Apply the blood or
control solution
sample only after
“Apply Blood,”
“Alternate Site,” or
“Control Solution”
appears on the
display. Refer to
pages 30–31 for
information on
sample application.
WARNING
See Owner's Booklet
Error 2
Meter or strip problem
Please repeat test
WARNING
See Owner's Booklet
Error 3
Meter not ready
Please repeat test
DCO_06158108A_OUS_OB_GB 5/6/03 1:25 PM Page 49