6 editing the stations, Editing the stations – LEICA ASP300 User Manual
Page 46

46
Instruction Manual V 1.3 – 12/2003
Fig. 23
↓
→
→
Adding / changing reagents
Highlight the desired station.
•
Press REAGENT NAME.
•
The SELECT REAGENT entry window pops up.
Only those reagents that are compat-
ible with the respective station will be
available for selection.
•
Highlight the desired reagent and press OK
to confirm.
5.2.6 Editing the stations
Fig. 23
shows all stations / allocated reagents cur-
rently defined in the instrument.
Allocating reagents
•
Only processing reagents (except paraffin)
can be allocated to stations 1 – 10.
•
Only a cleaning reagent can be allocated to
station 11.
•
Only a cleaning alcohol can be allocated to
station 12.
•
Only cleaning water/detergent can be allo-
cated to stations 13 and 13-ext.
•
Only paraffin wax can be allocated to the
wax baths.