Operation – LEICA ASP300 User Manual
Page 35

35
Leica ASP300
Program Options
Selection of the options on how programs are carried out.
•
STATION ALLOCATION:
By age or Sequential
By age
-
RMS is activated, reagents will automatically be
used in order of increasing cleanliness.
Sequential -
reagents will be used in sequential order of stations.
•
WAX BATH ORDER:
Auto or 1.; 2. ; 3.
Wax bath order can be activated only when STATION ALLOCATION
is set to Sequential.
•
PROMPT FOR NUM BLOCKS :
Enabled or Disabled
Enabled-
When starting a program, the actual number of
blocks must be entered.
This is recommended with the RMS being enabled.
Disabled -
Number of blocks can be entered optionally (not
mandatory).
Date / Time
Be sure to verify that the date and time entered do actually correspond
to the local time/date.
Press the buttons to access the entry windows where time/date can be
corrected, if necessary.
Instrument
Press WAX BATH SET TEMP (°C) to access the wax bath temperature
entry window. Select the highest temperature compatible with the type
of paraffin used as paraffin temperature will drop slightly once the paraf-
fin is in the retort.
5.