Setting up the instrument, Connecting the instrument to mains – LEICA ASP300 User Manual
Page 20

20
Instruction manual V 1.3 – 12/2003
4.
Setting up the instrument
•
Connect mains cable (6) to mains power supply socket (1) (see
Fig. 3
).
•
Jumper cable (3) connects the mains power supply output (2) to the
electronics module input (4).
•
Switch ON/OFF switch (5) ON.
•
Switch ON/STOP switch at the right of the instrument ON.
Fig. 3
1
2
6
3
4
5
Electrical
connections at the
rear panel of the
instrument
Cables
connected
correctly
Important:!
Specification for connection (2) is as follows:
100 - 120 V or 230 - 240 V, max. 200 VA.
Connecting the instrument to mains
The instrument MUST be connected to an earthed mains power
outlet socket.
It is recommended that the ASP300 be plugged into a wall socket
that has Ground Fault Circuit Interruption (GFCI) protection - as
an additional electrical safeguard.
Make sure to use the appropriate mains cable for the local volt-
age supply (wall outlet).