Instruction Manual V 1.3 – 12/2003
Select station
Press key symbol
Fig. 12
→
5.
Operation
Remote drain
•
On the screen, select reagent bottle no. 3
and the bulk container (Remote).
The press STATION DRAIN.
•
Check for leaks.
Remote fill
•
On the screen, select reagent bottle no. 3
and the bulk container (Remote).
The press STATION REFILL.
•
The remote fill procedure takes less than 3
minutes.
•
Check for leaks.
Fill / drain reagents
In FAVORITES, press MENU. In MENU FUNC-
TIONS, press SMART SCREEN.
The SMART SCREEN (
Fig. 12
) pops up.