4 technical data, Technical data, Instrument components and specifications – LEICA ASP300 User Manual
Page 15
15
Leica ASP300
3.4 Technical data
Nominal voltage:
Two factory-preset voltage settings
(not user-adjustable):
100 to 120 V + 1 0% or
230 to 240 V + 1 0%
Nominal frequency:
50 to 60 Hz
Main fuses:
2 fast-blowing melting fuses, 20 x 5 mm, UL-listed
Order no.
•
100 to 120 V; F 10 A 250 VAC
69411002
•
230 to 240 V; F 5 A 250 VAC
69411000
Nominal power:
1,000 VA
Dimensions (W x D x H), in mm:
595 x 680 x 1,325
Dry weight, unpacked:
approx. 140 kg
Weight, packed:
220 kg
Operating temperature range:
15 °C to 35 °C
Relative humidity:
10% to 80% - non-condensing
IEC 1010 classification:
Protective Class 1
Pollution Degree 2
Overvoltage Installation Category II:
•
800 V impulse (120 V systems)
•
1500 V impulse (240 V systems)
Maximum operating altitude:
maximum 2,500 m
Local / remote alarm relays:
30 V DC, maximum 2 A
2 connections:
both single voltage-free, change-over contacts
(can be operated as both normally-open and
normally-closed connections)
Paraffin wax baths
Number of baths:
3
Volume (liters):
4.3 l per bath
Melt time:
approx. 10 hrs
Temperature:
40 to 65 °C
Temperature accuracy:
+ 1 °C
3.
Instrument components and specifications