3 installing the accessories, Installing the accessories, Setting up the instrument – LEICA ASP300 User Manual
Page 21

21
Leica ASP300
4.
Setting up the instrument
•
Unpack the activated carbon filter and in-
sert it into the instrument (see
Fig. 4
). Make
sure to insert the filter with the correct side
facing upwards. See label (6) at the front of
the filter - the arrow must point upwards.
•
Unpack the magnetic stirrer and insert into
the retort.
•
Install the console flap (covers floppy drive
and ports).
l
Prior to filling the bottles with reagent for the
first time, check the connections. If neces-
sary, tighten manually.
6
Fig. 4
4.3 Installing the accessories
Remote fill / drain hose
•
Connect the remote fill / drain hose as
shown in
Fig. 5
.
•
Important!
When inserting the hose into the fill/drain
connection (item 7 in
Fig. 5
) the connecting
device (8) of the hose must lock with an au-
dible click.
Fig. 5
8
7
7