5.1.2 Changing the wax
Draining the wax
•
Connect the paraffin drain hose (2) as shown
in
Fig. 13
and insert the loose end of the hose
into a remote collecting container.
Fig. 13
•
Select the wax bath to be drained and the
remote container (‘Remote’).
Then press STATION DRAIN.
•
Several queries will be displayed each of
which has to be confirmed individually.
1
2
From the FAVORITES screen, press SMART
SCREEN (see previous chapter).
•
Make sure the hose end is securely kept in
the remote container until both the drain has
been completed and the hose has been
purged with pressurized air.
→
+
→