23. Carefully vacuum the areas between the lamps (both upper and

lower), then use a Staticide wipe to thoroughly clean the glass dust
plate between the lamps.

24. Clean the imaging guides thoroughly with a Staticide wipe.

IMPORTANT: Staticide wipes contain isopropanol which can cause eye

irritation and dry skin. Wash your hands with soap and
water after performing maintenance procedures. Refer to
the MSDS for more information.

25. Reinstall the upper and lower imaging guides. Handle the imaging

guides carefully as to not put fingerprints on the guides.

26. Close the pod.

27. Reinstall the output tray.

28. Power on the scanner.