beautypg.com

1 measuring ranges, 2 connection type, 3 range monitoring – Pilz PSSu E S 2AI RTD User Manual

Page 13: Measuring ranges, Connection type, Range monitoring

background image

Function description

Operating Manual PSSu E S 2AI RTD(­T)
22017­EN­06

13

}

Via a fieldbus with the modular device description file 
With a modular device description file it is possible to configure all the values.

}

Via the USB port of the head module, using the PSSuniversal Configurator
All the values can be configured using the PSSuniversal Assistant. Any configuration
via the USB port will overwrite the configuration made via the fieldbus.

If a module has been configured via the head module's USB port, it is locked and cannot be
overwritten by the fieldbus. This lock can be deactivated again in the PSSuniversal Config­
urator.

The following options exist for configuring the module in system environment B:

}

Via the head module's USB port with PAS4000.

4.2.1

Measuring ranges

The input measuring ranges may be configured per input:

}

Pt100 (Default)

}

Pt200

}

Pt500

}

Pt1000

}

Ni100

}

Ni120

}

Ni1000

}

0 ... 4000 Ohm (resistance measurement)

The temperature ranges for these measuring ranges can be found in the technical details.

4.2.2

Connection type

The measuring circuits for the temperature measurements may be configured per input:

}

2­wire measurement

}

3­wire measurement (default)

}

Connection for 4­wire sensor

The module will detect an open circuit on the measuring lines. Should an open circuit occur,
the module will output 7FFF as the measured value. If resistance measurement is con­
figured, the module will output FFFF in the case of an open circuit. With a 4­wire sensor
connection, an open circuit on an individual measuring line will not be detected; con­
sequently, the measurement result will be corrupted.

4.2.3

Range monitoring

You can deactivate range monitoring for each input. Range monitoring refers to the meas­
uring range of the sensor which is configured for the input. This option is activated in the
default configuration. The module writes the result of range monitoring as follows:

}

System environment A:

In the status byte (see 

PSSu assignment in system environment A [

 20]

)

}

System environment B:

This manual is related to the following products: