8.3 Device-Specific Commands

: SUORce : JITTer: RANGe?

Command type

Response

Function

Restriction

Example of use

Transmission sequential command

=

Makes a query of the transmission jitter generation range.

Effective when Jitter is generated.

Restrictions on setting apply depending on the Bit Rate.

Bit Rate

Ul Range

lOG

3,200, 80, 0.5 UI

5G

1,600, 40, 0.5 UI

2.5 G

800, 20. 0.5 UI

> : SOURce : JITTer ; RANGe?

< UI05

: SUORce : JITTer : AMPLitude?

Command type

Response

Function

Example of use

Sequential command

=

Performs reading of the value of Tx JITTer AMPLitude.

< ; SOURce : JITTer ; AMPLitude?

> 1.51

8-13