Section 8 Device Message Details
:
STATUS : OPERation : INSTrument: PTRansition?
Command type
Sequential command
Response
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Function
Makes a query of the details of the transition filter (a change in the positive direction) of the
INSTrument status register.
Example of use
When the bit of measurement completion is set at PTRansition.
> : STATUS : OPERation : INSTrument : PTRansition?
<2
:
STATUS : OPERation : INSTrument: NTRansition
Command type
Sequential command
Parameter
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Function
Sets the transition filter (a change in the negative direction) of the INSTrument status register.
Set at 1.
Example of use
Sets the bit of measurement completion at NTRansition.
> : STATUS ; OPERation ; INSTrument : NTRansition 2
: STATUS : OPERation : INSTrument ; NTRansition?
Command type
Response
Function
Example of use
Sequential command
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Makes a query of the details of the transition filter (a change in the negative direction) of the
INSTrument status register.
When the bit of measurement completion is set at NTRansition.
> : STATUS : OPERation : INSTrument : NTRansition?
<