B.1 initializing bus – Anritsu MP1777A User Manual
Page 119
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

B.1 Initializing Bus
IFC:
Initializes functions of all the devices connected to the GPIB bus line, when the GPIB option is
mounted.
Initialization of the interface function refers to an action of canceling the status of the interface function
of the device set by the controller and returning it to the initial status. Each function marked with a
check mark (vi) in the table is initialized. The function marked with a delta (A) means that the function
will be partially initialized. This command has no effects on the operation status of the device.
Table B-1 Changes in Status of Interface Functions by IFC Message
Appendix B Details of Initialization Command Functions
No.
Function
Cord
Initialization by IFC
I
Source handshake
SH
V
2
Acceptor handshake
AH
V
3
Talker or extension talker
T orTE
4
Listener or extension listener
L or LT
V
5
Service request
SR
A
6
Remote local
RL
7
Parallel port
PP
8
Device clear
DC
9
Device trigger
DT
10
Controlle
C
V
The following table shows changes in status of the device by the IFC message.
Table B-2 Changes in Status of Device by IFC Message
IFC Message
Changes in Status of Device
Talker/Listener
All the talkers and listeners switch into the idle state (TIDS and LIDS) within 100 ps.
Controller
Unless the controller is active (SACS: System control ACtive State), the controller switches
into the idle state (CIDS: Control IDle State) within 100 ps.
Device of service request
Although the state where the device is sending an SRQ message to the controller (state
where the SRQ line is set by the device to the LOW level) is not canceled, the message
cancels the state where the controller have all the devices under the system bus in the serial
pole mode.
Device in remote mode
Devices currently in the remote mode will not be released from the remote mode by the IFC
message.
B-2