Ionode IJ pH V5.1 User Manual
Page 28
Ionode IJ Instruction Manual
28
Symptom Possible
Causes
Remedy
Low Slope
(<94%)
Contaminated
Buffers
Replace buffers
Membrane
not Clean
See 4.6
Troubleshooting
Drift
Membrane
Aged
Allow more time for
stabilisation, or
replace
Electrode
Polarised
Replace electrode
Slow
Response
Membrane
not Clean
See 4.6
Troubleshooting
Drift
Membrane
Aged
Allow more time for
stabilisation, or
replace
Incorrect
Sample
Reading
after
successful
calibration
Ground Loop.
Often Occurs
in Process
Systems
Verify by removing
the sample from its
environment and
measuring in a glass
beaker. May require
special circuitry.
Junction
Blocked or
unclean
See 4.6
Troubleshooting
Drift
Displays
pH7 for all
Buffers
Electrical
Short
Check connector
Internal Short Replace electrode
Displays
pH4-5 for
all Buffers
Membrane or
Stem
Cracked
Replace electrode