beautypg.com

Ionode IJ ORP User Manual

Ij 64, Ij au, Ij ag

background image

Warranty

IJ electrodes have a pro-rata warranty of 12 months from
date of purchase. Any electrode found to be faulty due to
manufacture within this time will be replaced.
Ionode reserves the right to limit or modify product warranty
if it is deemed that the electrode has been used in unsuitable
applications.
IJ electrodes have an expected lifetime of 2-4 years in ideal
samples at room temperature. This may be reduced in
chemically aggressive or abrasive samples and at high
temperatures.
Please visit www.ionode.com.au to register your purchase as
soon as you receive your electrode.

User maintenance and troubleshooting

All specifi cations and values are subject to change without notice. © 2009

Intermediate Junction Series

Short-Form

Operators Manual

Download a detailed manual at

www.ionode.com.au

IJ

64

ORP Electrode (Pt tip)

IJ

Au

ORP Electrode (gold tip)

IJ

Ag

Silver Billet Electrode

Symptom Possible cause

Remedy

Drift

Junction blocked

Remove sleeve. Clean
junction or ground glass
stem

Sensor tip not
clean

Remove sleeve
Clean sensor tip

Reading stable
in standards, but
not in samples

See ORP theory in
comprehensive manual
available on-line

Electrode polarised

Replace electrode

Noisy

Poor connection
to meter

Check connection

Junction not
immersed fully

Lower electrode into
solution below junction

Insuffi cient
electrolyte

Refi ll electrolyte

Not within
tolerance
(±20mV)

Contaminated
standard

Replace standard

Correlation to
wrong reference
electrode

Calculate what standard
would read if using a
saturated KCl reference

Sensor not clean

Remove sleeve. Clean
sensor wire

Electrode
polarised

Isolate electrode

Displays
0mV for all
solutions

Electrical short

Check connector

Internal short

Replace electrode

Reading
out of
range

Noise

Refer above

Unreliable
titration
results
with IJAg

pH too high

Acidify with nitric acid to
approx pH2-3

Contamination
from electrolyte

Use 1M potassium nitrate
in sleeve

Slow ORP
response

Non reversible
ORP reaction

Stir sample

Parameter

Operating Range

mV range

-2000 to +2000 mV

Accuracy

± 10mV

Temperature range

0 – 60ºC

Reference type

Double Junction Ag/AgCl/sat KCl

Sensor materials

Platinum wire (IJ64)
Silver billet (IJAg)
Gold wire (IJAu)

Body and sleeve

Polypropylene

Overall length

150mm

Barrel diameter

12mm

Cable length

1m standard, longer to order.
Maximum 20m

Connector

BNC standard, others on request

All specifi cations and values are subject to change without notice. © 2009