Ionode IJ pH User Manual

Ij 44, Intermediate junction series, Operators manual

background image

Warranty

IJ electrodes have a pro-rata warranty of 12 months from
date of purchase. Any electrode found to be faulty due to
manufacture within this time will be replaced.
Ionode reserves the right to limit or modify product warranty
if it is deemed that the electrode has been used in unsuitable
applications.
IJ electrodes have an expected lifetime of 2-4 years in ideal
samples at room temperature. This may be reduced in
chemically aggressive or abrasive samples and at high
temperatures.
Please visit www.ionode.com.au to register your purchase as
soon as you receive your electrode.

User maintenance and troubleshooting

All specifi cations and values are subject to change without notice. © 2009

Intermediate Junction Series

Short-Form

Operators Manual

Download a detailed manual at

www.ionode.com.au

IJ

44

A/C

pH electrodes

Symptom Possible cause

Remedy

Drift

Junction blocked

Remove sleeve. Clean
glass stem and sleeve

Membrane not
clean

Remove sleeve.
Clean glass membrane

Membrane aged/
damaged

Allow longer measuring
time or replace

Temperature
changes

Stabilise temperature

Noisy

Bubble in
membrane

Flick electrode
downward

Poor connection

Check connection

Junction not
immersed fully

Lower electrode into
solution below junction

Insuffi cient
electrolyte

Refi ll electrolyte

Low Slope
(<95%)

Contaminated
buffers

Replace buffers

Membrane not
clean

Clean membrane

Membrane aged

Allow more time for
stabilisation, or replace

Electrode
polarised

Isolate sample
electrically

Slow
Response

Membrane not
clean

Clean membrane

Membrane aged

Allow more time for
stabilisation, or replace

Incorrect
Reading
after Cal

Junction blocked
or unclean

Clean junction

Reads pH7
all Buffers

Electrical short

Check connector

Internal short

Replace electrode

Reads
pH4-5 all
Buffers

Membrane or
stem cracked

Replace electrode

Large
Offset
(>0.5pH)

Insuffi cient sleeve
electrolyte

Refi ll electrolyte

Junction blocked/
unclean

Clean junction

Electrode
polarised

Isolate sample
electrically

Parameter

Operating Range

pH range

0-14 pH (C glass) 0-12 pH (A glass)

Zero potential point

pH 7.0 ± 0.5

Temperature range

0 – 60ºC

Reference type

Double Junction Ag/AgCl/sat KCl

Body and sleeve

Polypropylene

Overall length

150mm

Barrel diameter

12mm

Cable length

1m standard, longer to order.
Maximum 20m

Connector

BNC standard, others on request