4.6. Troubleshooting

A fault may develop in either the meter or
electrode. If a number of electrodes and/or
meters are available, interchange to isolate the
fault.

Table 4: Problem Indication

Symptom

Possible
Causes

Remedy

Drift

Junction
Blocked

Remove sleeve.
Clean glass stem
and sleeve. See
4.4.1

Cleaning

Membrane
not Clean

Remove sleeve.
Clean glass
membrane. See
4.4.1

Cleaning

Membrane
Aged or
Damaged
(e.g. by
Abrasion)

Allow longer
measuring time or
replace.

Electrode
Polarised

Replace electrode

Noisy

Bubble in
Membrane

Flick electrode
downward.

Poor
Connection to
Meter

Check connection

Junction not
Immersed
properly

Lower electrode into
solution below level
of junction (see

Figure 6

).

Insufficient
Electrolyte

See 3.

Preparation