4.4.1. Cleaning

As

pH

and

ORP

are

surface

based

measurements, the cleanliness of the sensor
and

junction

is

critical

for

accurate

measurement. For new electrodes, drift, poor
slope and long response times are often due to
an unclean sensor and/or junction.

Clean the electrode periodically depending on
the application.

Remove the sleeve (see 3

. Preparation

). Clean

the membrane, ground glass stem and sleeve
with solvents, detergents, or acid. DO NOT
USE ABRASIVE MATERIALS!

a) Alkaline samples and scale: soak the

membrane in approximately 1M HCl for an
hour. Wash well with tap water and soak
before use.

b) Grease and oil: wipe the membrane and

ground stem with cotton or tissue soaked in
acetone followed by methylated spirits.
Wash with deionised water and soak before
use.

c) Solids and organics: wipe the membrane

and junction with cotton or tissue soaked in
a mild non-alkaline detergent (e.g. Palmolive
™). Wash with deionised water and soak
before use.

d) Strongly absorbed impurities and chemically

bonded substances such as silicones can be
removed with non-

abrasive Jif ™ detergent.

This removes a fine layer of contaminated
pH glass without damaging the membrane.
Jif ™ can also be used for routine cleaning
(e.g. every three months)

. Use Jif ™

undiluted on a soft cloth and scrub the
membrane well. Wash well and soak.

e) Protein contamination: soak the membrane

in 1% pepsin in 0.1M HCl for 15 minutes
followed by rinsing with deionised water, and
soak in 20% KCl before use.