Principles of operation, Safety features and quality control – ITC ProTime Microcoagulation System User Manual
Page 5

5
PRINCIPLES OF OPERATION 
The ProTime Microcoagulation System measures the PT using fibrin clot formation and detection. The 
ProTime cuvette is a self-contained, micro volume reaction cell constructed of precision-molded plastic. 
There are two user options within the ProTime Microcoagulation System: the standard ProTime cuvette and 
the ProTime3 cuvette. These cuvettes differ from each other in the amount of blood that needs to be 
collected and tested. 
The standard ProTime cuvette has five micro-channels, which contain the dried reagents required to 
perform triplicate testing of the PT assay and two levels of controls. The ProTime3 cuvette has three 
functional micro-channels. Two micro-channels perform the controls, and one micro-channel performs the 
PT test. The standard ProTime uses the Tenderlett Plus device for performing the fingerstick, and it is 
designed to hold 65 µL of blood (approximately 3 drops) needed to fill all five micro-channels. The 
ProTime3 uses the Tenderlett Plus LV (low volume) device for performing the fingerstick, and it collects 27 
µL of blood (approximately 1 large drop) needed to fill the three micro-channels of the ProTime3 cuvette. 
The instrument draws the precise volume of blood into the micro-channels of each cuvette, which contain 
thromboplastin and other reagents. An array of LEDs detects the motion of sample/reagent mixtures as they 
move through a precision restriction in each channel. The blood is pumped back and forth until a clot 
forms, obstructing the channel and slowing the flow of blood. The instrument detects the clot when the 
blood movement decreases below a predetermined rate. 
SAFETY FEATURES AND QUALITY CONTROL 
Each ProTime and ProTime3 cuvette has two channels for performing the two levels of control which work 
simultaneous with the channel(s) used for testing the PT assay. The built-in safety features of the 
instrument and integral reagent controls work together to ensure that the instrument and reagent systems 
are working properly and that the test procedure was performed correctly. Two levels of quality control are 
performed with each and every test. 
Calibration 
The ProTime instrument and cuvettes are pre-calibrated. No additional calibration is required. 
Reagents 
ProTime cuvettes are pre-loaded with dried thromboplastin, stabilizers and buffers. The thromboplastin 
has a high sensitivity, measured as an ISI near 1.0. Each cuvette performs the prothrombin time test and, 
in addition, has one channel for a Level I control and one channel for a Level II control. The controls 
consist of purified plasma-extracted coagulation factors and anticoagulants. 
