Main menu options – ITC ProTime Microcoagulation System User Manual
Page 16

16
Screen Display
Source of Error
Method of Control
CUVETTE EXPIRED 
 
Expired cuvette
Check to make sure cuvettes have not 
expired. The expiration date is located 
on the packaging and alongside the 
barcode on the cuvette. 
BARCODE ERROR 
REMOVE CUVETTE 
TRY AGAIN 
Instrument cannot read 
barcode accurately 
Visually inspect barcode. If scratched, 
discard cuvette. If dirty, wipe clean. If 
barcode is good, review correct 
procedure and repeat test. 
BATTERY ERROR 
TURN OFF 
TRY AGAIN 
Power supply battery error
Repeat the test. If it occurs again, 
replace the battery. If problem 
persists, call ITC Technical Support. 
INSTRUMENT ERROR 
TURN OFF 
TRY AGAIN 
Instrument set up, data log 
or communication error 
Repeat the test. If problem persists, 
call ITC Technical Support. 
CHARGE ERROR 
TURN OFF 
TRY AGAIN 
Power supply error
Check the AC/DC power module, 
repeat the test. If problem persists, 
call ITC Technical Support. 
TEMP ERROR 
TURN OFF 
TRY AGAIN 
Temperature not in range
Check for proper operating 
temperature. Repeat the test. If 
problem persists, call ITC Technical 
Support. 
PHOTO ERROR 
TURN OFF 
TRY AGAIN 
LED blocked or other photo 
system error 
Repeat the test. If problem persists, 
call ITC Technical Support. 
MAIN MENU OPTIONS 
The options presented in the MAIN MENU are: 
Each of these options will lead to sub menus within the selection. The following buttons are used to navigate 
the menu: 
•
The button is used to move the highlight bar to select the option.
•
The button is used to select the option that is highlighted.
