ITC ProTime Microcoagulation System User Manual
Page 37

37
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ProTime® Microcoagulation System is intended for use in the electromagnetic environment specified 
below. The customer or the user of the ProTime® Microcoagulation System should assure that it is used in 
such an environment. 
Emissions Test
Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The ProTime
®
Microcoagulation System uses RF energy only for
its internal function. Therefore, its RF emissions are very low 
and are not likely to cause any interference in nearby electronic 
equipment. 
RF Emissions
CISPR 11
Class B
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/ 
Flicker Emissions 
IEC 61000-3-3
Complies
The ProTime
®
Microcoagulation System is suitable for use in all
establishments, including domestic establishments and those 
directly connected to the public low-voltage power supply 
network that supplies buildings used for domestic purposes. 