Absolute mode – In-Situ Tube 300R Telemetry System User Manual
Page 41

800-446-7488
41
www.in-situ.com
When a Warning is triggered, the measurement is stored in the log memory even if the
Tube is not designated to save that reading. When an Alarm is triggered, the reading is
stored and an alarm message is sent (if the corresponding alarm has been enabled).
Alarm Mode is set in the Probe and Set-Point Configuration Menu, under the
Parameters Monitoring tab.
Absolute Mode
In Absolute Mode, alarms and warnings occur when a measurement value is less than
the minimum threshold value or greater than the maximum threshold value.
Minimum Level Value
Maximum Level
Value
Measurement Value
Result
1
15
16
ALARM
1
15
1
OK
1
15
15
OK
1
15
0
ALARM
-2
10
1
OK
See also other documents in the category In-Situ Water equipment:
- TROLL Shield Guard for Aqua TROLL 100 Instruments (1 page)
- Aqua TROLL 400 Modbus and SDI-12 (79 pages)
- TROLL 9500 Low-Flow System with RDO Sensor (2 pages)
- Conductivity/Level/Temp Meter (2 pages)
- Oil/Water Interface Meter (5 pages)
- Water Level Tape 100 (2 pages)
- Aquaculture Aerator Controller (22 pages)
- Aquaculture Buoy (40 pages)
- Aquaculture Buoy (34 pages)
- Aqua TROLL 100 (94 pages)
- Aqua TROLL 100 (2 pages)
- Level TROLL 400 (83 pages)
- Level TROLL 400 (4 pages)
- Level TROLL 300 (84 pages)
- Rugged TROLL 100 (46 pages)
- PXD-261 (4 pages)
- Rugged TROLL (4 pages)
- Virtual HERMIT Pump Test Kit (2 pages)
- Aqua TROLL 400 (41 pages)
- RDO PRO-X Process Optical Dissolved Oxygen Sensor (29 pages)
- RDO PRO Process Optical Dissolved Oxygen Sensor (26 pages)
- smarTROLL Multiparameter Handheld (71 pages)
- smarTROLL Multiparameter Handheld (2 pages)
- smarTROLL RDO Handheld (38 pages)
- smarTROLL RDO Handheld (2 pages)
- TROLL 9500 (164 pages)
- TROLL 9500 (2 pages)
- smarTROLL Low-Flow Sampling System (2 pages)
- Con TROLL PRO (29 pages)
- Con TROLL PRO (32 pages)
- Con TROLL PRO (55 pages)
- Con TROLL PRO (40 pages)
- Con TROLL PRO (2 pages)
- Con TROLL PRO (2 pages)
- RuggedReader (81 pages)
- TROLL Link Telemetry System (50 pages)
- Blue RuggedReader (2 pages)
- Telemetry System: Cube 300 (2 pages)
- Telemetry System: Cube 300S (2 pages)
- Telemetry System: Tube 300R (2 pages)
- Telemetry System: Tube 300S (2 pages)
- TROLL Link 100 (4 pages)
- TROLL Link 101 (2 pages)