In-Situ Con TROLL PRO Operators Quick Start Guide (purchased after 10 May 2012) User Manual
Con troll, Pro module
0085700 rev. 002 10/2013
Con TROLL
®
PRO Module
_________________________________________
Keypad and Display Window
Use the Left and Right function keys to
choose the command shown on the screen
directly above each key.
Use the Up/Left arrow key for scrolling
through options on the main display area of
the screen.
Use the Down/Right arrow key to move down
or across through lists of options.
__________________________________________
Getting Started
Before making any wiring connections between
probes and the module, make sure that the probes
are set to their factory-default communication
settings. If you are using new probes, settings will be
correct. If you are using pre-existing probes, use
Win-Situ software to reset the communication
settings.
After the Con TROLL PRO module has been
installed by a qualified professional using the
Installation Manual, you can begin operating the
keypad. The DC model display turns on after you
push any button. The AC model display is always on
when the unit is connected to power.
1. Select a language using the arrow keys. Select
Enter to accept the language.
2. Adjust the Contrast level using the arrow keys.
Select Enter to accept the contrast.
3. Use table on the back of the Quick Start Guide
to determine your UTC time. Use the Up/Left
arrow key to enter a positive or negative value.
Push the Down/Right arrow key to move the
appropriate numeric position. Use the Up/Left
arrow key to enter the correct numeric value.
Repeat for additional digits as needed. Push
Enter to accept the time.
4. Enter the date using the Up/Left arrow key.
Select Enter to accept.
5. Enter the time using the Up/Left arrow key.
Select Enter to accept.
6. Enter the location using the Up/Left arrow key.
Select Enter to accept.
7. The
“Add Probe A” screen appears. Select
Enter to configure the probe and sensors.
8. The
“Add Probe B” screen appears. If there is
no Probe B, the module will display that no
probe is connected. Select Enter to configure
this sensor if it exists, or to proceed to the next
screen if it does not.
9. Parameters are labeled A, B, and C.
References to Probe C refer to the Con TROLL
PRO itself.
__________________________________________
Main Screen
The Main Screen shows two parameters from Probe
A, one parameter from Probe B and one parameter
from the Con TROLL PRO. You can display four
parameters at a time.
__________________________________________
View Option
1. Push the View button. An arrow will point to the
first menu item. Push the View button again to
move to the next menu item.
2. With the arrow pointing to a menu item, use the
Up and Down arrow buttons to view the
available parameters on the probes and the
Con TROLL PRO.
__________________________________________
Administrator and User Password Access
The User can access the Calibrate, Log Data, Hold
Outputs, and Diagnostics menus. The Administrator
can access all functions, most notably the Settings
menu. If there is no administrator code set, the User
becomes the default Administrator and can access all
functions. To remove pass code-protected access,
enter “0000” for all codes.
1. From the main screen, select Menu, Settings,
Security Code, Administrator.
2. Set a 4-digit code for the Administrator using
the arrow keys. Push Enter.
3. Push User. Set a 4-digit code for the User using
the arrow keys. Push Enter.
__________________________________________
Adding or Replacing a Probe
If you want to replace a probe with the same model
and type of probe (i.e., Level TROLL
®
500 with a
Level TROLL 500 or Aqua TROLL
®
400 with an Aqua
TROLL 400) you can do this directly and maintain
existing settings.
Important: Download all data before adding or
removing a probe. Data logged to the Con TROLL
PRO will be deleted when you replace a probe.
Danger: See the Safety section on page 2 of
this Quick Start Guide before proceeding.
1. Refer to the Con TROLL PRO Installation
Manual, Chapter 5, Input/Output Connections
for directions on safely disconnecting the
existing probe and replacing it with the same
model of sensor.
2. Replace the sensor as instructed in the Con
TROLL PRO Installation Manual.
3. Select Main Menu, Settings, Probes, the probe
type, Replace Probe, Enter. The controller will
configure the probe. Hold down Back to return
to the Main menu.
__________________________________________
Logging Mode and Data Download
Con TROLL PRO Models AC-L and DC-L have
logging capabilities.
1. From the main screen, select Menu, Data Log.
2. Select the probe and parameter you want to
view.
3. Use the arrow buttons to scroll through the
data.
4. Data can be downloaded to a RuggedReader
®
Handheld PC or a computer using a TROLL
Com or Bluetooth. See the Con TROLL PRO
Operator’s Manual for more information.