Cd indicators – Ohaus CD-33 Indicator User Manual
Page 41
41
CD Indicators
Command
Command
Command
Command
Command
Character
Character
Character
Character
Character
Description
Description
Description
Description
Description
?
?
?
?
?
Print current mode: g, kg, custm, %, #
nnnA
nnnA
nnnA
nnnA
nnnA
Set Auto Print feature to "nnnn".
nnnn=0 Turns feature OFF
nnnn=S Output on stability
nnnn=C Output is continuous
nnnn=1-3600 Sets Auto Print interval
C
C
C
C
C
Begin span calibration.
F
F
F
F
F
Print current function.
xl
xl
xl
xl
xl
Set Averaging Filter to "x", where x = 0 to 3.
If LFT, level 0 to 1.
0=minimum level
1=
2=
3= maximum level
L
L
L
L
L
Begin linearity calibration.
xM
xM
xM
xM
xM
Places terminal in mode "x", where x = 1 to 4.
If unit is not already enabled, command returns ES.
1=grams
2=kilograms
3= pounds
4= custom units
P
P
P
P
P
Print display data.
xSL
xSL
xSL
xSL
xSL
Set stable data only printing TO "X", where x = 0 to 3.
If LFT, x=0 or 1.
0=.5d
1= 1d
2= 2d
3=5d
T
T
T
T
T
Same effect as pressing Tare Key.
N
N
N
N
N
Shows copyright notice.
V
V
V
V
V
Print software version number.
EscV
EscV
EscV
EscV
EscV
Print scale ID
xZ
xZ
xZ
xZ
xZ
Set Auto Zero to "x", where x=0 to 3.
If LFT, Auto Zero level from 0 to 1.
0=off
1=0.5d
2=1d
3=3d
x%
x%
x%
x%
x%
Set % reference weight. Uses real number "x" as current Percent Reference Weight.
Weight must be entered in Calibration Units.
x#
x#
x#
x#
x#
Set PC reference weight. Uses real number "x" as current Parts reference Weight. Weight
must be entered in Calibration Units.
EscR
EscR
EscR
EscR
EscR
Resets setup and print menus to factory defaults.
CAUTION: This will reset RS232 configuration.
#
#
#
#
#
Print current Parts Count Reference Weight in the most recently used Units.
%
%
%
%
%
Print current Percent Reference Weight in the most recently used Units.
xF
xF
xF
xF
xF
Set current function to "x", where x=0 to 4.
If mode is not already enabled, command returns ES.
0=None (normal weigh)
RS232 COMMAND TABLE
4.7 Printing Data (Cont.)
4.7.2 Output Formats (Cont.)