4 good manufacturing practices (gmp) set, Cd indicators procedure – Ohaus CD-33 Indicator User Manual
Page 19
19
CD Indicators
Procedure
To select any of the items in the GMP Set menu, proceed
as follows:
3.4 Good Manufacturing Practices (GMP) Set
Good Manufacturing Practices (GMP) Set submenu allows the selection of: Time, Scale Identification Number, User
Identification Number, Project Number, Difference and Name data to be printed. When the selected items are set to
ON, these items are displayed. The default setting is OFF. When an external printer is used, and all items are set ON
and the Indicator is calibrated, the printer will print out calibration data for audit trail purposes and will indicate date,
and time. (It should be noted that the User ID number and Project number must be entered in the GMP Data submenu
before printed data is available).
• Press the Setup button, SCALE is displayed.
• Press
button until GMP SET is displayed.
• Press Enter button to save setting.
• Press
or
button until either TIME, SCALE
ID, USER NO, PROJ NO, DIFF, NAME, LOCK or
EXIT is displayed. Use the next four steps to
turn each entry ON or OFF and to advance to next
menu item.
• Press Enter button to enter submenu.
• Press
or
button and select either ON or
OFF.
• Press Enter button to save setting.
• Press
or
button to continue or EXIT.
• Press Enter button to save setting.