3 good manufacturing practices (gmp) data, Procedure – Ohaus CD-33 Indicator User Manual
Page 18
18
CD Indicators
3.3 Good Manufacturing Practices (GMP) Data
The GMP Data submenu enables the storage of a user identification number (7 digits), a project number (7 digits), and
a Scale ID number (7 digits). When entered into the Indicator, the identification number, project number and Scale ID
are available when printing providing they are turned on in the GMP Set submenu. A lock setting is also available
which locks the settings.
Procedure
To select any of the items in the GMP Data menu,
proceed as follows:
• Press the Setup button, SCALE is displayed.
• Press
or
button until GMP DATA is dis-
played.
• Press Enter button, USER NO is displayed.
• Press Enter button to continue.
• Press
or
buttons or numerical
buttons and enter a 7 digit number for the user USER
NO number.
• Press Enter button to save setting, PROJ NO is dis-
played.
• Press Enter button to continue.
• Press
or
buttons or numerical
buttons and enter a 7 digit number for the project
number.
• Press Enter button to save setting, SCALE ID is dis-
played.
• Press Enter button to continue.
• Press
or
buttons or numerical
buttons and enter a 7 digit number for the balance ID
number.
• Press Enter button to continue, LOCK is displayed.
• Press Enter button. Using arrow buttons, select ON or
OFF. When Lock Switch jumper is in place, LOCK
OFF = data can be changed. When Lock Switch
jumper is removed, LOCK ON= data is locked and
cannot be changed.
• Press Enter button, EXIT is displayed, press Enter
button, Indicator saves settings and returns to weigh
mode.