5 calibration gmp printout, Cd indicators 4.1.5 calibration gmp printout, Span calibration printout – Ohaus CD-33 Indicator User Manual
Page 33: Linearity calibration printout, Calibration test printout

33
CD Indicators
4.1.5 Calibration GMP Printout
If any option in the GMP Set Menu is turned On, GMP automatically prints data after calibration is completed.
- - - - - SPAN CAL - - - - - -
12/01/99
1:10:00 PM
Scale Id 1234
Cal:
40kg
Old:
39.99kg
Dif:
0.01kg
Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - LIN CAL - - - - - -
12/01/99
1:20:00 PM
Scale Id 1234
Cal:
40kg
Old:
39.99kg
Dif:
0.01kg
Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - CAL TEST - - - - - -
12/01/99
1:30:00 PM
Scale Id 1234
Cal:
10kg
Act:
10kg
Dif:
0.00kg
Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
Span Calibration Printout
When performing Span calibration with all GMP options
turned on, a printout is automatically made after the
calibration is completed.
Linearity Calibration Printout
When performing a Linearity calibration with all GMP
options turned on, a printout is automatically made after
the calibration is completed.
Calibration Test Printout
When performing a Calibration Test with all GMP options
turned on, a printout is automatically made after the
calibration is completed.