4 parts counting, Cd indicators 4.4 parts counting, Procedure – Ohaus CD-33 Indicator User Manual
Page 37

37
CD Indicators
4.4 Parts Counting
Parts Counting is
enabled only when Count is turned ON in the Mode submenu under Setup and selected with the
Mode button. In the parts counting mode, the Indicator displays the quantity of parts you place on the pan. Since the
Indicator determines the quantity based on the average weight of a single part, all parts must be reasonably uniform in
weight.
Procedure
• Press the Mode button.
• Press
or
button until COUNT is displayed.
• Press Enter button to save setting, PUT>PAN 10 PC
is displayed (default setting), Indicator will retain last
sample size saved.
• Press Tare if taring is required. Container can be
tared anytime during operation; container on scale,
press Tare button. When Tare was entered, display
always shows NET weight (G/N/T button disabled).
• Press
or
button and select sample size.
Sample size is 5 to 1000 pieces.
• Place sample size on the pan.
• Press Enter button to continue, display indicates
WORKING.
Indicator displays the reference weight of an individual
piece part for 5 seconds and then displays the total
number of pieces on the pan.
• Remove the sample and place parts to be counted on
the pan. Indicator displays number of pieces.
• To return to weighing mode, press MODE button,
COUNT is displayed.
• Press
button until WEIGH is displayed, then
press Enter button. Indicator is now in a weighing
mode.
(Tare was entered)