Cd indicators 2.6 scale base setup (cont.), Procedure – Ohaus CD-33 Indicator User Manual
Page 13
13
CD Indicators
2.6 Scale Base Setup (Cont.)
Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the
indicator are compatible with the scale base. Alternate displays are shown in this procedure to save space.
Procedure
• Press the Setup button, SCALE is displayed.
ZERO
Zero specifies the percentage of full capacity load (2%,
18% or 100%) that may be cleared by pressing >0<.
Limits may be required in certain applications.
NOTE: 2% Zero capacity is used for hopper scales or
other large scales where accidental zero would lose the
current weight. Bench and counter scales normally use
100% Zero capacity. Legal for trade applications use
2%.
• Press Enter button, ZERO is displayed.
• Press Enter button, ZERO 2% (default) is displayed.
• Press
and select either 2%, 18% or 100%.
• Press Enter button, SAVED is momentarily displayed
then CALUNIT is displayed.
• Press Enter button, CALUNIT kg is displayed.
• Press
and select either kg or lb, then press
Enter button, SAVED is momentarily displayed then
GRAD.
• Press Enter button, GRAD .01 is displayed.
• Press
and select desired setting from .0005
to 100 for scale graduation.
• Press Enter button, SAVED is momentarily displayed
and FULL SC is displayed.
• Press Enter button, FULL SC with a 7 digit number is
displayed, with the first digit flashing. Illustration
indicates a 40 kg scale setting.
• Press
and or
and enter the
desired scale base capacity when using Model CD-31.
The numerical keypad can be used for Model CD-33.
Remember that this number represents either pounds
or kilograms, whatever was previously set for cal unit.
• Press Enter button, SAVED is momentarily displayed
followed by LOCK.