Calibration path – Health O Meter PROPLUS 2000KL User Manual
Page 26

25
CALIBRATION PATH
The calibration is performed in Kg or Lb, according to the units of measure used upon entering 
into the calibration path. 
Operator Action
Display
1. Press ON/OFF button and immediately press
and hold HOLD/RELEASE button for 3-4 
seconds 
Calibration
Process
2. Press
ENTER
Enter load weight
400.0
3. Using
cd keys and/or the keypad, set the
calibration load to at least 200 lbs (for best 
results and greater accuracy, use a 400 lb 
weight). Press ENTER 
Zero calibration 
Please clear the scale 
4. Please clear the stretcher and press ENTER
Zero calibration
Please wait
5. Please wait without touching or shaking the
scale until the zero calibration process is 
finished 
Weight calibration
Put: XXX.X
6. Load the stretcher with the required weight and
press ENTER
Weight calibration
Please wait
7. Please wait without touching or shaking the
scale until the calibration process is finished
Calibration 
factor: X.XXXXX 
8. Press
ENTER
Calibration 
Please clear the scale 
9. Please remove weight from the stretcher
Rebooting 
Please wait 
10. Please wait until the scale resumes normal
operation
Health O Meter 
Pro Plus 
EXPLODED VIEW OF SCALE
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52