Amprobe AMB-110 Insulation-Resistance-Tester User Manual
Page 39
34
Withstanding voltage DC
• DC test voltage:
• Voltage value: Any value within 500V and 10kV.
• Accuracy: -0 / +10 % + 20 V.
Display range Test voltage (V)
Resolution
Accuracy
0 - 9999 V
1 V
(3 % of reading + 3 V)
10 kV
0,1 kV
(3 % of reading)
Leakage current
Display range Itrigg (mA)
Resolution
Accuracy
0.000 - 0.009
1 A
(3 % of reading + 3 digits)
0.01 - 5,50
10 A
(3 % of reading)
Voltage
Voltage AC or DC
Display range External Voltage (V)
Resolution
Accuracy
0 - 600
1 V
(3 % of reading + 4 V)
Frequency of external voltage
Display range (Hz)
Resolution
Accuracy
0 and 45 - 65
0.1 Hz
0.2 Hz
Note
• for frequency between 0 and 45 Hz displayed <45 Hz
• for frequency over 65 Hz displayed >65 Hz
Input resistance: 3 M
e
10 %
Capacitance
Measuring range C: 50 F*
Display range C
Resolution
Accuracy
0.0 - 99.9 nF
0.1 nF
(5 % of reading + 2 digits)
100 - 999 nF
1 nF
1.00 - 50.00 F
10 nF
*Full-scale value of capacitance is defined according to the following equation:
CFS = 10 F * U
test
[kV]