Amprobe AMB-110 Insulation-Resistance-Tester User Manual
Page 37
![background image](https://www.manualsdir.com/files/864272/content/doc037.png)
32
Measuring range Riso: 0.12 Me up to 10 Te*)
Display range Riso
Resolution
Accuracy
0 - 999 ke
1 ke
(5 % of reading + 3 digits)
1.00 - 9.99 Me
10 ke
10.0 - 99.9 Me
100 ke
100 - 999 Me
1 Me
1.00 - 9.99 Ge
10 Me
10.0 - 99.9 Ge
100 Me
100 - 999 Ge
1 Ge
1.00 -10.00 Te
10 Ge
(15 % of reading + 3 digits)
*Full-scale value of insulation resistance is defined according the following equation:
R
FS
= 1 Ge * U
test
[V]
DC test voltage:
• Voltage value: Any value within 500 V and 10 kV, steps by 25 V.
• Accuracy: -0 / +10 % + 20 V.
• Output power: 10 W max.
Display range Test voltage (V)
Resolution
Accuracy
0 - 9999 V
1 V
(3 % of reading + 3 V)
10 kV
0,1 kV
(3 % of reading)
Current:
Display range I (mA)
Resolution
Accuracy
1.00 - 5.50 mA
10 A
(5 % of reading + 0.05 nA)
100 - 999 A
1 A
10.0 - 99.9 A
100 nA
1.00 - 9.99 A
10 nA
100 - 999 nA
1 nA
10.0 - 99.9 nA
0.1 nA
0.00 - 9.99 nA
0.01 nA
Noise current rejection (resistive load)
Filter option
Maximum current @ 50Hz (mA r.m.s).
Fil0
1.5
Fil1
2.5
Fil2
4.5
Fil3
5