MSO1000Z/DS1000Z Quick Guide

7

Probe Compensation

When the probes are used for the first time, you should compensate the probes to

match the input channels of the oscilloscope. Non-compensated or poorly

compensated probes may cause measurement inaccuracy or error. The probe

compensation procedures are as follows:

1. Perform steps 1, 2, 3 and 4 of "Function Inspection".

2. Check the displayed waveforms and compare them with the following figures.

Figure 9 Probe Compensation

3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment

hole on the probe until the displayed waveform is as the “Perfectly compensated”

in the figure above.

Over compensated Perfectly compensated Under compensated