MSO1000Z/DS1000Z Quick Guide

IX

Working Environment

Temperature

Operating: 0℃ to +50℃
Non-operating: -40℃ to +70℃

Humidity

0℃ to +30℃: ≤95% relative humidity
+30℃ to +40℃: ≤75% relative humidity
+40℃ to +50℃: ≤45% relative humility

WARNING

To avoid short circuit inside the instrument or electric shock, please do not operate in

humid environment.

Altitude
Operating: below 3 km
Non-operating: below 15 km

Installation (Overvoltage) Category
This product is powered by mains conforming to installation (overvoltage) category II.

WARNING

Make sure that no overvoltage (such as that caused by thunderbolt) can reach the

product, or else the operator might expose to danger of electric shock.

Installation (Overvoltage) Category Definitions

Installation (overvoltage) category I refers to signal level which is applicable to equipment

measurement terminals connected to the source circuit. In these terminals, precautions are done

to limit the transient voltage to the corresponding low level.

Installation (overvoltage) category II refers to the local power distribution level which is applicable

to equipment connected to the AC line

(AC power).

Pollution Degree

Degree 2

Pollution Degree Definitions

Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no

influence. For example: a clean room or air-conditioned office environment.

Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary

conductivity caused by condensation may occur. For example: general indoor environment.

Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which

becomes conductive due to condensation which is expected. For example: Sheltered outdoor

environment.

Pollution degree 4: Pollution that generates persistent conductivity through conductive dust, rain,

or snow. For example: outdoor locations.