beautypg.com

Specifications, Electrical characteristics – RIGOL DS6000 Series User Manual

Page 40

background image

RIGOL

8

RP1002C User’s Guide

Specifications


These characteristics apply to an adjusted RP1002C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with
the temperature at 10℃-30℃ and the humidity at 0-80%.

Electrical Characteristics

Current Range

50mV/A, 500mV/A

DC Accuracy, typical

500mV/A: ±3% ±20mA
(20mA to 14A peak range)
50mV/A: ±4% ±200mA
(200mA to 100A peak range)
50mV/A: ±15% maximum
(100A peak to 140A peak range)

Gain versus frequency,
typical

See Appendix 1 Gain versus
Frequency at 1 A Peak (Typical)

Max. Working Current

See Voltage and Current Ratings

Max. Working Voltage

See Voltage and Current Ratings

Max. Floating Voltage

See Voltage and Current Ratings

Frequency Range

DC to 1MHz (-3dB)

Battery Type and Life,
typical

9V NEDA 1604A, IEC 6LR61
8 hours (1 each)

DC Signal Linearity, typical See Appendix 3 DC Signal Linearity in

50mV/A Range (Typical)

Phase versus frequency,
typical

See Appendix 4 Phase versus
Frequency at 1 A Peak (Typical)




This manual is related to the following products: