2

RP1002C User’s Guide

Control/Indicator Description

Current flow symbol: The arrow shows
the positive direction of current flow
.

Zero adjustment: After connecting to
power and fixing the current probe to a
fixed position, rotate the knob to adjust
the probe output tothe proper DC offset. It
may also be used to offset a DC signal
component. Zeroing is not needed for AC
measurements unless your instrument
cannot isolate a DC component (if
present).

Range switch: Slide the switch from OFF
to either the 50 mV/A or 500mV/A range.
When 50 mV/A or 500mV/A is selected,
the probe is turned on, and the green
battery indicator (ON) lights.

Power Supply indicator: The green
power supply indicator lights when the
probe is turned on. For more information,
please refer to To Install the Battery.

Overload indicator: The red overload
indicator lights and flicks continuously if
the measured signal is greater than the
maximum range of the selected range
capacity.