Basic operation – RIGOL DS6000 Series User Manual
Page 33
RIGOL
RP1002C User’s Guide
1
Basic Operation
Before using the current probe, the battery or specified AC power
adapter must be installed. For the battery installation instructions,
refer to To Install the Battery.
CAUTION
Do not clamp the probe onto circuits with voltages
greater than 600 VAC. Otherwise, personal injury or
damage to the probe may result.
CAUTION
Always connect the RP1002C current probe output to the
instrument before clamping onto the circuit under test.
Connection between the Current Probe and
Oscilloscope
1 Set the coupling mode of the oscilloscope to DC and connect
the BNC interface of the RP1002C current probe to the
oscilloscope input via dual-BNC coaxial cable.
2 Move the Range switch to the 50 mV/A or 500 mV/A position
to turn on the RP1002C current probe. At this point, the green
battery indicator lights.
3 Use the zero adjustment knob to adjust the probe to the
proper DC offset.