Perform a Blower Door test on the building and

record the EqLA at 50 Pa.

Install cardboard in upper part of doorway with a

20 x 20 inch hole cut in it.

Perform a second Blower Door test on the

building, record the EqLA at 50 Pa.

Subtract the first result from the second result

and the value should be 400 sq. in. (+/-10%).

Speed control is handled automatically with FanTestic
software, for complete automation.

Control fan speed with software

Field check system monthly

Alternatively, use a Verification Plate in an optional

double hole Door Panel, or use the optional Flex Duct

with a 400 sq. in. hole in a plate on the end.

Page 4 of 4

Field check gauge weekly

Check gauge operation and check for blocked, leaking or

pinched tubes weekly, and anytime results are in question.

To perform the gauge check, you will need the gauge

and Umbilical.

Set [Time Average] to 5 seconds in [Settings].

Tap [Channel B] and select“Pressure: Pa”.

Connect the red tube between the red and

yellow ports.

Repeat between different ports with each of the

tubes you use for testing.

If readings on Channel A and Channel B are within

2% and don’t drop rapidly, the tube is not blocked

or leaking and the gauge is correct.

Checking your gauge

and tubes regularly will

eliminate a common

source of error in

readings.