beautypg.com

Electromagnetic compatibility – Precision Digital PD663 User Manual

Page 10

background image

PD663 Loop-Powered Meter

Instruction Manual

10

Electromagnetic Compatibility

EMISSIONS

EN 61326:2006
Safety requirements for measurement, control, and
laboratory use – Industrial
Group 1 Class A ISM emissions requirements

Radiated
Emissions

Class A

IMMUNITY

EN 61326:2006
Safety requirements for measurement, control, and
laboratory use

ESD

±4 kV contact,
±8 kV air

RFI – Amplitude
Modulated

80-1000 MHz @ 10 V/m,
1.4-2.0 GHz @ 3 V/m,
2.0-2.7 GHz @ 1 V/m,
80% AM (1 kHz)

EFT

±2 kV DC mains, ±1 kV other

Telco Surge

±1 kV

CRFI

3 V, 0.15-80 MHz, 1 kHz 80% AM