Appendix c: non-standard toxic gas sensors, Specifications – RKI Instruments Eagle Series User Manual
Page 64

57 • Appendix C: Non-Standard Toxic Gas Sensors
Eagle Instruction Manual
WARNING: The Eagle’s alarms are initiated by the DISPLAY reading
not the FACTORED reading. If you are monitoring for
toluene as in the above example and the low alarm is set
for 10% LEL, the Eagle will initiate a low alarm at 7%
LEL toluene (display reading of 10% LEL).
Appendix C: Non-Standard Toxic Gas Sensors
Appendix C describes the Eagle’s non-standard, electrochemical toxic
gas sensors. This appendix also offers information for calibrating and
replacing non-standard toxic gas sensors. Non-standard electro-
chemical toxic gas sensors are toxic gas sensors other than CO or H
2
S.
Specifications
Table 8 lists specifications for the non-standard toxic gas sensors. The
alarm settings are user-adjustable (see “Updating the Alarm Point
Settings” on page 33).
Table 8: Non-Standard Toxic Gas Sensors Specifications
Target Gas
Range
Alarm 1
Alarm 2
TWA
Alarm
STEL
Alarm
Ammonia
0 to 75.0 ppm
12.0 ppm
25.0 ppm
25.0 ppm
35.0 ppm
Arsine
0 to 1.00 ppm
0.05 ppm
0.10 ppm
0.05 ppm
OFF
Arsine
0 to 200 ppb
1
50 ppb
100 ppb
50 ppb
OFF
Bromine*
0 to 1.00 ppm
0.05 ppm
0.10 ppm
0.10 ppm
OFF
Chlorine
0 to 3.00 ppm
0.50 ppm
1.50 ppm
0.50 ppm
1.00 ppm
Chlorine
Dioxide*
0 to 1.00 ppm
0.10 ppm
0.30 ppm
0.10 ppm
0.30 ppm
Diborane
0 - 300 ppb
1
50 ppb
100 ppb
100 ppb
OFF
Fluorine**
0 to 5.00 ppm
0.50 ppm
1.00 ppm
1.00 ppm
2.00 ppm
Hydrogen
Bromide*
0 to 9.00 ppm
1.00 ppm
2.00 ppm
OFF
3.00 ppm
Hydrogen
Chloride*
0 to 15.0 ppm
2.0 ppm
5.0 ppm
OFF
OFF
Hydrogen
Cyanide
0 to 30 ppm
3.0 ppm
5.0 ppm
OFF
4.7 ppm
Hydrogen
Fluoride**
0 to 9.00 ppm
1.00 ppm
2.00 ppm
1.00 ppm
2.00 ppm