Entering setup mode, Updating the battery type setting – RKI Instruments Eagle Series User Manual
Page 32

25 • Setup Mode
Eagle Instruction Manual
• To exit setup mode, from the main menu place the prompt next to
the last menu option, START MEASUREMENT, then press the
POWER/ENTER button. The Eagle begins its normal start-up
sequence.
Entering Setup Mode
WARNING: The Eagle does not detect gas or display readings while
in setup mode. The CAL/SETUP switch (SW2) must be in
the ON position to enter setup mode.
1. Take the Eagle to a non-hazardous location, and turn the power
off.
2. Press and hold down the AIR/▲ and SHIFT/▼ buttons, then press
the POWER/ENTER button. The main menu displays. It displays
four menu options at a time. Press the AIR/▲ or SHIFT/▼ button
to view additional menu options.
Updating the Battery Type Setting
This setting allows you to select between alkaline and Ni-Cd batteries.
The Eagle uses this setting to make sure adequate time is given
between the low battery warning and low battery alarm. This setting
has no effect on battery charging.
1. From the main menu, select the BATTERY TYPE menu option.
2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.
CAUTION:
This setting should always match the type of batteries (alkaline
or Ni-Cd) installed in the Eagle. If this setting does not match
the installed batteries, the time between low battery warning
and low battery alarm may be less than expected.
3. Press the POWER/ENTER button to enter the setting and return to
the main menu.