RKI Instruments Eagle Series User Manual
Page 35

Eagle Instruction Manual
Setup Mode • 28
setting. The fullscale setting flashes.
The maximum fullscale setting for the combustible gas channel is
50,000 PPM; the minimum setting is 1000 ppm. The default setting
is 50,000 ppm.
2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired
fullscale setting (see Table 5), then press the POWER/ENTER
button to enter the setting. The prompt flashes.
Returning to the Main Menu
1. Press the SHIFT/▼ button. The
ESCAPE
message displays. (Press
the AIR/▲ button to return to the previous screen.)
2. Press the POWER/ENTER button to save the new setting. The
OTHER GAS SET
message displays, then the Gas Combinations
menu displays.
3. To exit the Gas Combinations menu, press the SHIFT/▼ button
until the prompt is next to Channel 4, then press SHIFT/▼ again.
The
ESCAPE
message displays.
4. Press the POWER/ENTER button. The message
SAVING DATA
displays, then the main menu displays.
Table 5: Fullscale PPM Readings
Equivalent to 100% LEL
Target Gas
Fullscale Setting
Methane (CH
4
)
50,000 ppm
Hexane
11,000 ppm
Hydrogen
40,000 ppm
Pentane
15,000 ppm
Styrene
9,000 ppm
IPA
20,000 ppm
Isobutane
18,000 ppm
Propane
21,000 ppm
Propylene
20,000 ppm
Toluene
11,000 ppm
Ethane
30,000 ppm
Ethanol
33,000 ppm
Benzene
12,000 ppm