Updating the alarm latching setting, Updating the alarm silence setting – RKI Instruments Eagle Series User Manual
Page 42

35 • Setup Mode
Eagle Instruction Manual
2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.
3. Press the POWER/ENTER button to enter the setting and return to
the main menu.
Updating the Alarm Latching Setting
With Alarm Latching ON, the Eagle remains in alarm condition until
the alarm condition passes and the RESET/SILENCE is pressed.
With Alarm Latching OFF, the Eagle automatically resets its alarm
when the alarm condition passes.
1. From the main menu, select the ALARM LATCHING menu
option.
2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.
3. Press the POWER/ENTER button to enter the setting and return to
the main menu.
Updating the Alarm Silence Setting
NOTE:
This feature works only when Alarm Latching is turned on.
With Alarm Silence ON, pressing the RESET/SILENCE button
silences the buzzer when the Eagles goes into alarm. The LEDs
continue to flash, and the display continues to show the level of
alarm. When the gas concentration falls below the low alarm level,
pressing the RESET/SILENCE button turns off the LEDs and removes
the
ALM1
message.
With Alarm Silence OFF, you cannot silence the buzzer.
1. From the main menu, select the ALARM SILENCE menu option.
2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.
3. Press the POWER/ENTER button to enter the setting and return to
the main menu.