Updating the Display Increment Setting
1. Press the SHIFT/▼ button to place the prompt in the third line,
then press the POWER/ENTER button. The display increment
setting flashes.
The minimum display increment setting is 0.1 PPM; the maximum
display increment setting is 2.5 PPM.
2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired
display increment setting, then press the POWER/ENTER button
to enter the setting. The prompt flashes.
Returning to the Main Menu
1. Press the SHIFT/▼ button. The
ESCAPE
message displays. (Press
the AIR/▲ button to return to the previous screen.)
2. Press the POWER/ENTER button to save the new settings. The
OTHER GAS SET
message displays, then the Gas Combinations
menu displays.
3. To exit the Gas Combinations menu, press the SHIFT/▼ button
until the prompt is next to Channel 4, then press the SHIFT/▼
button again. The
ESCAPE
message displays.
4. Press the POWER/ENTER button. The message
SAVING DATA
displays, then the main menu displays.
Updating Combustible Gas Channel Units of Measure
This setting allows you to display the combustible gas reading in
percentage of LEL or percentage of volume. The detection range
remains the same. If 100% LEL equals 5% by volume, then fullscale on
the volumetric display is 5%.
1. From the main menu, select the LEL% OR VOL% (HC) menu
option.
2. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.
NOTE:
The data logging option logs all combustible gas readings in
LEL% regardless of this setting.
3. Press the POWER/ENTER button to enter the setting and return to
the main menu.
%