Page 9

background image

MODEL 5081-T

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

1.2.3 TRANSMITTER SPECIFICATIONS @ 25°C

Measured Range*: 50 to 2,000,000 µS/cm (see chart)

Accuracy: ± 1.0% of reading

Repeatability: ± 0.25% of reading

Stability: 0.25% of output range/month,

non-cumulative

Ambient Temperature Coefficient: ± 0.2% of FS/°C

Temperature Slope Adjustment: 0-5%/° C

% Concentration Ranges:

Sodium Hydroxide: 0 to 15%
Hydrochloric Acid: 0 to 16%
Sulfuric Acid: 0 to 25% and 96 to 99.7%

1.2.4 LOOP SPECIFICATIONS

Loop Accuracy: With a standard Model 228 or 225 sen-

sor with 20' cable, laboratory accuracy at 25°C can be as
good as ± 2% of reading and ± 50 µS/cm.

To achieve optimum performance, standardize the sen-
sor in the process at the conductivity and temperature of
interest.

Results under real process conditions, at different tem-
peratures, or using other sensors may differ from above.

RTD accuracy: Utilizing a perfect 100 Ohm RTD after 1

point temperature standardization, temperature reading
can be as good as ± 0.5°C.

RECOMMENDED SENSORS:

Model 222

Flow-Through

Model 225 Clean-In-Place (CIP)
Model 226 Submersion/Insertion
Model 228 Submersion/Insertion/Retractable
Model 242* Flow-Through

*no I.S. approval for loops of 5081-T with 242-06 or 242-08

1.3 HAZARDOUS LOCATION APPROVAL

Intrinsic Safety:

Class I, II, III, Div. 1
Groups A-G

T4

Tamb = 70°C

Exia Entity

Class I, Groups A-D

Class II, Groups E-G

Class III

T4

Tamb = 70°C

ATEX

1180

II 1 G
Baseefa03ATEX0399
EEx ia IIC T4
Tamb = -20°C to +65°C

Non-Incendive:

Class I, Div. 2, Groups A-D

Dust Ignition Proof

Class II & III, Div. 1, Groups E-G

NEMA 4X Enclosure

Class I, Div. 2, Groups A-D

Suitable for

Class II, Div. 2, Groups E-G

T4 Tamb = 70°C

Explosion-Proof:

Class I, Div. 1, Groups B-D

Class II, Div. 1, Groups E-G

Class III, Div. 1

Class I, Groups B-D

Class II, Groups E-G

Class III

Tamb = 65°C max

3

RECOMMENDED RANGES FOR TOROIDAL SENSORS

Conductivity Sensor

Model Number

226

228

225

222 (1in.)

222 (2 in.)

242

Nominal Cell Constant

1.0

3.0

3.0

6.0

4.0

*

Minimum Conductivity (

µµS/cm)

50

200

200

500

500

100*

Maximum Conductivity (

µµS/cm)

1,000,000

2,000,000

2,000,000

2,000,000

2,000,000

2,000,000*

* Model 242 values depend on sensor configuration and wiring.