1 accessing the calibrate menu – Emerson 5081-T User Manual
Page 52

46
MODEL 5081-T
SECTION 6.0
START-UP AND CALIBRATION
SECTION 6.0
START-UP AND CALIBRATION
6.1 ACCESSING THE CALIBRATE MENU
The “Calibrate” menu is used to calibrate the transmitter to known temperature and conductivity values. This menu also con-
tains the temperature calibration operation to establish the temperature slope.
Figure 6-1 illustrates the relationship between the Calibrate Menu and its sub-menus. Each sub-menu leads to a series of
prompts that are used for calibration.